15
8. Specifying Measuring Conditions
When using this tester to measure moisture or solid content, it is necessary to first specify the conditions
(e.g., drying temperature or measuring mode) of the measurement. The settings to be specified are de-
scribed below.
It is also possible to save sets of measuring conditions (i.e., drying temperature, measuring mode, and bias).
8-1 Selecting Temperature Display Mode
The temperature display that is used for drying temperature setting and temperature indication on the
main unit display can be selected from the following two modes.
[ST (Sample position Temperature)] mode
The temperature at the sample position is used for setting and displaying temperatures.
Sample position temperature is converted from the temperature detected by the thermistor near the
heater. The default setting is this mode.
TT (Thermistor Temperature)] mode
The temperature detected by the thermistor near the heater is used for setting and displaying
temperatures.
How to select temperature display mode
Key(s) used
Display
Operating instructions
From the weight display, press
SELECT
key first and while holding it
down, press
CONDITION
key and keep pressing both keys for about 5
seconds.
Currently selected temperature display mode is indicated as
either 'TT' or 'ST' at the M display part.
Every time
SELECT
key is pressed, the display is switched between
'TT' and 'ST'. Press
ENTER
kkey while the desired temperature
display mode is shown.
It returns to the weight display upon completion of setting
temperature display mode.
Step
1
2
3
4
SELECT
CONDITION
ENTER
SELECT
The temperature display mode selected here is effective for all measurement modes.
Indication of currently set mode:
In the display, the °C symbol
blinks
when the current setting is 'ST' mode and when the current
setting is 'TT' mode, the °C symbol is illuminated without blinking.
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Содержание MOC-120H
Страница 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...