62
SCAIME Manual : NU-eNod4C-E-0418-196712-F.docx
Outputs 1-4 timings
In Checkweigher, allows modifying timings on outputs with functions « Remote control » (activation time),
« Result within tolerances » and « Result out of tolerances » (activation time and delay).
Parameters Analog output (IO+ version only)
Fonction
The allocation of the analog output is unique for both outputs (current and voltage)
Polarity
Bipolar / Unipolar
Analog outputs can be assigned to gross measurement copy.
Maximal level value is related to maximum capacity (MC)
parameter and works in mono-quadrant functioning. Bipolar
option can only be applied to gross measurement copy. When
this option is activated, the lowest value of current and voltage
levels corresponds to -MC and the highest value to +MC.
Voltage Output
Voltage output might be set either 0-5V or 0-10V.
Current Output
Current output to 4-20 mA, 0-24 mA, 0-20 mA or 4-20 mA alarm 3.6 mA.
Analog output remote (Page 2) with remote control function