Analog output affectation function is common to both
current
and
voltage
output and might be assigned to followings:
•
None:
analog outputs have no function.
•
Gross measurement:
analog outputs can be assigned to gross measurement copy. Maximal level value is
related to
Maximum Capacity
parameter and works in mono-quadrant functioning. Bi-quadrant option can
only be applied to gross measurement copy. When this option is activated, the lowest value of current and
voltage levels corresponds to -
MC
and the highest value to +
MC
.
•
Net measurement:
analog outputs can be assigned to net measurement copy. Maximal value is related to
Maximum Capacity
parameter and works in mono-quadrant functioning only. The highest value of current
and voltage levels corresponds to +
MC
in only one quadrant.
•
Level on request:
analog outputs are driven by master requests through the
external value to control analog
output
variable (in 0.01% of the full scale of current or voltage analog outputs).
When analog output is assigned to “Gross measurement” or “Net measurement” its value jumps to a special error value
when the internal alarm flag described in
“Weighing diagnosis” § in the MEASUREMENT AND STATUS § is activated. This
allows to warn about defection of the measurement chain.
Содержание eNod4-C BOX
Страница 8: ...8 SCAIME Manual NU eNod4C E 0418 196712 F docx 3 GENERAL CHARACTERISTICS 3 1 Dimensions ...
Страница 9: ...9 SCAIME Manual NU eNod4C E 0418 196712 F docx ...
Страница 14: ...14 SCAIME Manual NU eNod4C E 0418 196712 F docx 4 CONNECTIONS ...
Страница 64: ...64 SCAIME Manual NU eNod4C E 0418 196712 F docx Parameter Possible Value Description Metrology Parameters ...