72
SCAIME Manual : NU-eNod4C-E-0418-196712-F.docx
12.2
Checkweigher setup menu (eNod4-C)
•
Detailed explanations on checkweigher functioning and parameters could be found into eNod4-C manuals
•
The following setup menu allows modifying the parameters related to checkweigher mode.
Parameter
Possible Value
Description
Dynamic Zero parameters
Dyn. Zero time
0
–
6553 ms
The ‘Dynamic zero’ function is the acquisition of a zero without
stability criterion, obtained by averaging measurements during the
specified time.
Automatic Dyn.
Zero
[Yes], [No]
Automatic zero correction allows to dynamically track the Zero
evolution. When enabled, an average Zero value is continuously
calculated, using the ‘Dyn Zero time’ and ‘Zero Correction range’
parameters.
Zero correction
range
0
–
255d