OPERATING INSTRUCTIONS
52
4618 - 021
This display will occur if the electrical signal produced
by the product passing through the aperture was
insufficient to guarantee the accuracy of the setting of
the phase control.
This display will occur if no product was passed
through the detector when requested.
This display may occur for one of the following
reasons.
1) The detector is sensing a signal when there should
be no product passing through the aperture.
Repeat the automatic setup and ensure that no
product is passed through the detector when
instructed, and that no product or metal is close to
the aperture and influencing the detector.
2) The automatic setup has requested the user to
'Pass product through aperture ....' and no product
has been sensed within the required period.
Repeat the automatic setup and ensure that the
product is passed through the aperture when
requested.
3) The automatic setup has not been able to setup
correctly due to the size of the product signal.
Switch 'Power drive' to LOW (page 45) and repeat
the automatic setup.
Содержание Signature
Страница 16: ......
Страница 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Страница 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Страница 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Страница 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Страница 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Страница 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Страница 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...