OPERATING INSTRUCTIONS
108
4618-021
RCU Fault Reporting.
The Module will display one of two RCU fault
conditions.
The fault displays will be shown from the running
mode display only, and will not prevent the Module
from operating. To draw attention to the problem the
buzzer will be driven.
Reject Fault
.
This fault display indicates that the RCU has signalled
a reject fault condition.
Assuming the RC extension window time has been
set correctly, the fault condition could be due to either
or both of the following;
a)
The Reject Mechanism has failed to operate or
failed to operate at the correct time.
b)
The RC Sensor has not signalled the RCU to
confirm reject operation.
The detector will clear the display when the RCU is
reset.
Input/output Fault
This display indicates that a fault has occurred in the
input output circuitry from the Module to the RCU, this
could be a connection or I.C fault within the Detector
Head or Module. It is more likely to be one of the
following;
a)
The RCU cable connection from the Detector
Head to the P/S Connection PCB is not
connected correctly.
b)
The RCU board is not in situ.
The Module will not clear the display until the Detector
Head is switched off.
Содержание Signature
Страница 16: ......
Страница 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Страница 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Страница 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Страница 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Страница 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Страница 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Страница 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...