OPERATING INSTRUCTIONS
48
4618 - 021
AUTOMATIC SETUP
Introduction
The automatic setup feature may be used to
automatically set the following;
•
Sensitivity
•
Phase
•
Frequency (see Note below)
•
Product
Signal
Cancellation
The automatic setup feature is designed to make
setting up your metal detector as easy as possible by
requesting the operator to pass the product through
the aperture at specific times and NOT requiring the
operator to adjust any of the controls.
Before using the automatic setup feature ensure that
a)
If a pack sensor is fitted that the pack sensor has
been set up correctly.
b)
That the speed of the product has been set-up
correctly.
c)
The sensitivity limit has been set correctly.
d)
If the application is an 'Inverse detect' application
that the 'Inverse detect' feature has been enabled
before starting automatic setup.
The automatic setup feature can NOT be used to set
up for products on a continuously running product
line.
The automatic setup feature will correctly set up for
products that exhibit or do not exhibit a product effect.
Normally the operator will have to pass the product
through the aperture about four times (see the section
below entitled “Notes on passing product through the
detector” At worst the operator will be requested to
pass the product through the aperture ten times.
If a pack sensor is to be used with the detector, set up
the pack sensor details before starting automatic
setup - for certain products the use of the pack sensor
will speed up the operation of automatic setup.
The automatic setup supports ‘Inverse detect’
applications. If required select the ‘Inverse detect’
feature ON for the current product number prior to
starting automatic setup.
Note
1) Automatic setup will attempt to setup the detector
in the currently selected Operating Frequency
setting. However there are two conditions where
automatic setup will change the currently selected
'Operating Frequency’ setting, these are;
a) If 'Operating Frequency’ is set to HIGH,
automatic setup will change this to LOW under
large product signal conditions.
b) If 'Operating Frequency’ is set to LOW,
automatic setup will change this to HIGH under
dry or very small product signal conditions.
NOTE
In a minority of cases following automatic setup it may
be necessary to manually trim the phase and/or
sensitivity to optimise performance. Refer to the
earlier Sensitivity and Phase display for details on
how to do this.
NOTE
Remember to record all settings on the Settings
Sheets at the rear of the manual.
Содержание Signature
Страница 16: ......
Страница 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Страница 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Страница 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Страница 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Страница 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Страница 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Страница 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...