Manual
Double Sheet Detector R1000 series E20
with integrated fieldbus interface
Maintenance
B0046191 / Rev. 1.12
136
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
For a 4P control unit with 2-4 sensors:
1
Verify that no sheet is in front of the sensor resp. the sensors
2 Press
the
MENU
key, switch to
“change“
and confirm with the
ENTER
key
3
Enter the password
4 Switch
to
“program parameter“
and confirm with the
ENTER
key
5
Verify that no sheet is in front of the sensor resp. the sensors
6 Switch
to
“zero adjust“
and confirm with the
ENTER
key
7 Select
“zero adjust: separate“
or
“zero adjust: common“
7.1
If
“zero adjust: separate“
, then confirm with the
ENTER
key
7.1.1
The display shows:
“zero adjust S1: on“
or
“zero adjust S1: off“
If zero adjust is desired, then select
„on“
and confirm with the
ENTER
key
.
if zero adjust is not desired (or sensor is not used in program),
then
select
“off“
and confirm with the
ENTER
key
.
7.1.2
The display shows:
“zero adjust S2: on“
or
“zero adjust S2: off“
If zero adjust is desired, then select
„on“
and confirm with the
ENTER
key
.
if zero adjust is not desired (or sensor is not used in program),
then
select
“off“
and confirm with the
ENTER
key
.
Step 7.1.2 repeats for sensor 3 (S3) and sensor 4 (S4) if they were selected.
7.2
If
“zero adjust: common“
, then confirm with the
ENTER
key,
all connected sensors will become zero adjusted.
8
The display will show
“zero adjust: active“
, and then
“zero adjust: off“
9
Exit the programming with the
MENU
key.
10.2
Exchange of control unit
Note
When exchanging the control unit, pay attention to the Hardware versions
of the systems!
From Hardware version 5 (from system no. 103227, from 27.02.2006)
different measurement times apply. Please see section 3.2 „Sensor data“
and chapter 13 „Sensor data“.
Exchange with RPP software
The E20 permits backup of the equipment parameters with the provided RPP-software via the
RS232 interface.
Exchange without RPP software
The original system configuration must be restored after an equipment exchange. In this
conjunction the documentation form “11.3 System configuration“ in the chapter “11. Technical
data“, should be used during the new start-up. If this form is not available, the information must be
retrieved from the equipment documentation.