Remote Control Commands
R&S
®
SMC100A
320
Operating Manual 1411.4060.32 ─ 04
:TRIGger<hw>[:SWEep]:SOURce
<Source>
The command sets the trigger source for all sweeps.
The names of the parameters correspond directly to the various settings under manual
control. SCPI uses other names for the parameters; these names are also accepted by
the instrument. The SCPI names should be used if compatibility is an important consid-
eration.
An overview of the various names is given in the
Parameters:
<Source>
AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled con-
tinuously. As soon as one sweep is finished, the next sweep is
started.
SINGle|BUS
One complete sweep cycle is triggered by the GPIB com-
mands
:SOURce:SWEep:POWer|FREQuency:EXEC
,
TRIGger:PSWeep|FSWeep:IMMediate
or
*TRG
.
If
:SOURce:SWEep:POWer:MODE
is set to
STEP
, one step is exe-
cuted.
The mode has to be set to
AUTO
.
EXTernal
The sweep is triggered externally via the INST TRIG connector.
EAUTo
The sweep is triggered externally via the INST TRIG connector.
As soon as one sweep is finished, the next sweep is started. A
second trigger event stops the sweep at the current frequency, a
third trigger event starts the trigger at the start frequency, and so
on.
*RST:
SINGle
Example:
TRIG0:SOUR EXT
selects triggering with an external trigger. The trigger is input via
the INST TRIG connector.
Usage:
Setting only
7.17 UNIT Subsystem
The UNIT subsystem contains the commands specifying which units are valid if no unit
is indicated in a command. These settings are valid for the entire instrument.
UNIT Subsystem