Remote Control Commands
R&S
®
SMC100A
271
Operating Manual 1411.4060.32 ─ 04
Example:
PULM:SOUR INT
selects the internal pulse generator as source for pulse modulation
PULM:STAT ON
activates pulse modulation. The pulse generator and the output of
the signals at the PULSE VIDEO connector are automatically acti-
vated in addition.
PGEN:STAT OFF
deactivates the output of the pulse signal by the pulse generator
at the PULSE VIDEO connector. The pulse modulation of the RF
carrier must be activated with command
SOURce:PULM:STATe
.
Usage:
SCPI confirmed
7.12.9 SOURce:PHASe Subsystem
This subsystem contains the commands for adjusting the phase of the RF output signal
relative to a reference signal of the same frequency.
.................................................................................................271
................................................................................271
[:SOURce<hw>]:PHASe
<Phase>
The command specifies the phase variation relative to the current phase. The variation
can be specified in RADians.
Parameters:
<Phase>
float
Range:
-359.9 to 359.9
Increment: 0.1 deg
*RST:
0.0 deg
Example:
PHAS 2DEG
changes the phase by 2 degrees relative to the current phase.
PHAS:REF
adopts the set phase as the current phase.
Usage:
SCPI confirmed
[:SOURce<hw>]:PHASe:REFerence
The command adopts the phase set with
SOURce:PHASe:ADJust
as the current phase.
Example:
PHAS 0.1RAD
changes the phase by 0.1 RAD relative to the current phase.
PHAS:REF
adopts the set phase as the current phase.
Usage:
Event
SCPI confirmed
SOURce Subsystem