Remote Control Commands
R&S
®
SMC100A
305
Operating Manual 1411.4060.32 ─ 04
Example:
SYST:COMM:NETW:HOST ‘SIGGEN’
enters the individual computer name of the R&S Signal Generator.
Example:
SYST:COMM:NETW:COMM:SET ON
activates the change of the hostname.
:SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup
<Workgroup>
The command enters the individual workgroup name of the instrument.
Parameters:
<Workgroup>
string
Example:
SYST:COMM:NETW:COMM:WORK 'TEST_09'
enters the workgroup name 'TEST_09'.
:SYSTem:COMMunicate:NETWork:IPADdress:MODE
<Mode>
The command selects if the IP address is assigned automatically or manually. The entry
is only activated after sending command
SYST:COMM:NETW:IPAD:SET
.
Parameters:
<Mode>
AUTO | STATic
AUTO
The IP address is assigned automatically. The network used must
support automatic assignment of address (DHCP) in order to use
this function.
STATic
The IP address is assigned manually.
*RST:
AUTO
Example:
SYST:COMM:NETW:IPAD:MODE AUTO
the IP address is assigned automatically (DHCP).
SYST:COMM:NETW:IPAD:SET ON
activates DHCP.
:SYSTem:COMMunicate:NETWork:IPADdress
<Ipaddress>
The command enters the IP address. The entry is only activated after sending command
SYST:COMM:NETW:IPAD:SET
.
Parameters:
<Ipaddress>
string
Range:
0.0.0.0. to ff.ff.ff.ff
Example:
SYST:COMM:NETW:IPAD 7.8.9.10
enters the IP address of the instrument.
Example:
SYST:COMM:NETW:IPAD:SET
activates all changes of the IP address settings.
SYSTem Subsystem