Instrument Function
R&S
®
SMC100A
136
Operating Manual 1411.4060.32 ─ 04
"Single"
Sets a single sweep cycle. The sweep is triggered by the "Execute Sin-
gle Sweep" button.
If a different sweep mode was activated prior to the "Single" mode, the
current sweep is stopped. The "Single" sweep always starts at the start
frequency when triggered.
The "Reset Sweep" button resets the sweep to the start frequency.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC
"Step"
Sets a step-by-step sweep cycle.
If this mode is activated, the cursor moves to the value displayed for
"Current Freq". Any variation to the "Current Freq" value triggers a
sweep step. The step width is set below at entry field "Step Lin" or "Step
Log".
If a different sweep mode was activated prior to the "Step" mode, the
current sweep is stopped. The "Step" sweep starts at the current RF
frequency when triggered.
The "Reset Sweep" button resets the sweep to the start frequency.
Example:
SOUR:SWE:FREQ:MODE MAN
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:SPAC LIN
SOUR:SWE:FREQ:STEP:LIN 0.5E4
SOUR:FREQ:MAN 1GHz
The value entered with command
SOUR:SWE:FREQ:STEP:LIN|LOG
sets the step width.
The value entered with command
SOUR:FREQ:MAN
has no effect, the
command only triggers the next sweep step. However, the value has to
be in the currently set sweep range (start to stop). In remote control only
a step-by-step sweep from start to stop frequency is possible.
RF Signal - RF Block