Remote Control Commands
R&S
®
SMC100A
215
Operating Manual 1411.4060.32 ─ 04
This mode is recommended for preserving the display especially if the instrument is
exclusively operated via remote control.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
DISP:PSAV ON
activates screen saver mode.
7.6 FORMat Subsystem
The FORMat subsystem contains the commands which determine the format of the data
that the R&S SMC returns to the controller. This affects all query commands which return
a list of numerical data or block data. Reference is made to this in the descriptions of the
commands.
:FORMat:BORDer
<Border>
The command determines the sequence of bytes within a binary block. This only affects
blocks which use the IEEE754 format internally.
Parameters:
<Border>
NORMal | SWAPped
NORMal
The instrument expects (with setting commands) and sends (with
queries) the least significant byte of each IEEE754 floating-point
number first and the most significant byte last.
SWAPped
The instrumenet expects (with setting commands) and sends (with
queries) the most significant byte of each IEEE754 floating-point
number first and the least significant byte last.
*RST:
NORMal
Example:
FORM:BORD SWAP
the data is transferred with the most significant bit first.
:FORMat[:DATA]
<Data>
The command determines the data format which the R&S Signal Generator uses to return
data. When data is transferred from the control computer to the instrument, the instrument
detects the data format automatically. In this case, the value set here is irrelevant.
FORMat Subsystem