Remote Control Commands
R&S
®
SMC100A
316
Operating Manual 1411.4060.32 ─ 04
7.16 TRIGger Subsystem
The
TRIGger
system contains the commands for selecting the trigger source for the RF
and LF sweep. The trigger input connectors are configured in the
SOURce:INPut
sub-
system.
The trigger system of the R&S SMC is a simplified implementation of the SCPI trigger
system. The
TRIGger
system differs from the SCPI system as follows:
●
No
INITiate
command; the instrument behaves as if
INITiate:CONTinuous
ON
were set.
●
Under
TRIGger
several sweep subsystems exist.
Other commands associated with the trigger system of the R&S SMC can be found in
the modulation and RF signal subsystems.
TRIGger<hw>
●
Suffix
TRIGger<1|2>
is not permitted
●
TRIGger0
activates the LF output.
Table 7-2: Cross-reference between the manual and remote control
R&S name
SCPI name
Command under manual control
AUTO
IMMediate
"Auto" mode
SINGle
BUS
"Single" mode.
EXTernal
EXTernal
"Ext Single" and "Ext Step" mode.
Use command
LFO:SWEep:MODE
to select between the two sweep
modes.
EAUTo
-
"Ext Start/Stop" mode.
...........................................................................................316
.................................................................................316
................................................................................317
....................................................................................317
................................................................................318
....................................................................................319
................................................................................319
....................................................................................320
TRIGger0[:SWEep]:SOURce
<Source>
:TRIGger<hw>:LFFSweep:SOURce
<Source>
The command sets the trigger source for the LF sweep. The trigger is triggered by the
command
:SOURce:LFOutput:SWEep[:FREQuency]EXECute
.
The names of the parameters correspond directly to the various settings under manual
control. SCPI uses other names for the parameters; these names are also accepted by
the instrument. The SCPI names should be used if compatibility is an important consid-
eration.
TRIGger Subsystem