Remote Control Commands
R&S
®
SMC100A
311
Operating Manual 1411.4060.32 ─ 04
Example:
SYST:ERR?
queries the oldest entry in the error queue.
Response: 0, 'no error'
No errors have occurred since the error queue was last read out.
Usage:
Query only
:SYSTem:IDENtification
<Identification>
Enables/disables selection of user defined "IDN String" and "OPT String" for the selected
instrument (
:SYST:LANG
).
Note
: While working in a emulation mode, the R&S SMC specific command set is disa-
bled, i.e. the SCPI command
SYST:IDEN
will be discarded.
Parameters:
<Identification>
AUTO | USER
*RST:
AUTO
Example:
SYST:IDEN USER
selects the user defined identification string.
SYST:IRES "Test Device"
defines the identification string 'test device'
*IDN?
Response: 'test device'
selects command set of the HP generator. The R&S SMC's com-
mand set is disabled.
To return to the SCPI command set of the R&S SMC, use the HP
command
EX
.
:SYSTem:IRESponse
<Iresponse>
The command defines the identification string for selection user defined (
SYST:IDEN
USER
).
Note
: While working in a emulation mode, the instrument's specific command set is dis-
abled, i.e. the SCPI command
SYST:IRES
will be discarded.
Parameters:
<Iresponse>
string
Example:
SYST:IDEN USER
selects an user-defined identification
SYST:IRES "Test Device"
defines the identification string 'test device'
*IDN?
Response: 'test device'