Remote Control Commands
R&S
®
SMC100A
318
Operating Manual 1411.4060.32 ─ 04
The names of the parameters correspond directly to the various settings under manual
control. SCPI uses other names for the parameters; these names are also accepted by
the instrument. The SCPI names should be used if compatibility is an important consid-
eration.
An overview of the various names is given in
Parameters:
<Source>
AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled con-
tinuously. As soon as one sweep is finished, the next sweep is
started.
SINGle|BUS
One complete sweep cycle is triggered by the GPIB commands
*TRG
. The mode has
to be set to
AUTO
(
:SOURce:SWEep:FREQuency:MODE AUTO
).
EXTernal
The sweep is triggered externally via the INST TRIG connector.
EAUTo
The sweep is triggered externally via the INST TRIG connector.
As soon as one sweep is finished, the next sweep is started. A
second trigger event stops the sweep at the current frequency, a
third trigger event starts the trigger at the start frequency, and so
on.
*RST:
SINGle
Example:
TRIG:FSW:SOUR EXT
selects triggering with an external trigger.
:TRIGger<hw>:PSWeep[:IMMediate]
The command immediately starts an RF level sweep.
The command is only effective for sweep mode "Single" (
SOURce:SWEep:POWer:MODE
AUTO
in combination with
TRIG:PSW:SOUR SING
).
The command corresponds to the manual control "Execute Single Sweep".
Example:
SWE:POW:MODE AUTO
selects the triggered sweep mode, i.e. a trigger is required to start
the sweep.
TRIG:PSW:SOUR AUTO
sets the Single trigger mode, i.e. a trigger starts a single sweep.
TRIG:PSW
starts a single RF level sweep.
Usage:
Event
TRIGger Subsystem