Remote Control Commands
R&S
®
SMC100A
240
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Offset>
float
Range:
-100.0 to 100.0
*RST:
0
Default unit: dB
Example:
SENS:POW:OFFS 10.0
sets a level offset of 10 dB
SENSe<ch>[:POWer]:OFFSet:STATe
<State>
The command activates the addition of the level offset to the measured value. The level
offset value is set with command
SENSe[:POWer]:OFFSet
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
SENS1:POW:OFFS 0.4dB
sets a level offset of 0.4 dB
SENS1:POW:OFFS:STAT ON
a level offset of 0.4 dB is added to the measured value.
SENSe<ch>[:POWer]:SNUMber?
The command queries the serial number of the sensor.
Return values:
<Snumber>
string
Example:
SENS:SNUM?
queries the serial number.
Usage:
Query only
SENSe<ch>[:POWer]:SOURce
<Source>
The command selects the signal source for the measurement.
Parameters:
<Source>
A | B | USER
*RST:
A
Example:
SENS:SOUR A
selects the RF signal as measurement source. The RF frequency
is used as the measurement frequency of the sensor and the cor-
responding correction factor is used. The level setting of the instru-
ment serves as reference level of the measurement.
SENSe, READ and INITiate Subsystems