Instrument Function
R&S
®
SMC100A
101
Operating Manual 1411.4060.32 ─ 04
The corresponding menu is available under "Save/Recall" in the "Setup" menu. The
instrument settings are saved in files which can be stored in data directories.
Additionally there are intermediate memories in which the current instrument setting can
be stored and then called up again by just pressing a key. This provides fast switching
between different instrument settings.
Only settings which differ from the preset values and configuration data for the operating
elements (e.g. window positions) are stored. As a result the files remain relatively small.
Furthermore, instrument settings can easily be transferred between different equipped
signal generators since the files contain only relevant information. When loaded, the ref-
erenced settings are implemented and all non-referenced parameters are set to the
associated preset values.
If list data is part of the instrument settings, e.g. a list of user correction data, a reference
to this list is stored, not the list itself. The list is reactivated when the associated settings
are loaded, but the list may have been modified or deleted in the meantime or may not
be available on a different instrument. If the list has been modified, the new entries will
be used. An error message appears if an attempt is made to access a non-existing list
or to activate settings which are not supported by the instrument.
●
Network settings and remote settings are not saved and restored.
●
Lists are stored and loaded in the appropriate menus. For example, the user correc-
tion data list is created and stored in the "User Correction" menu.
When loading an instrument setting, it is possible to select whether the current frequency
and level setting is to be retained or whether the stored settings are to be activated. It is
possible to delete stored instrument settings. A file can be copied by loading it with
"Recall" and then storing it under a new name.
Settings can be transferred easily between instruments with different equipment options
and/or firmware versions because only the settings which differ from the preset values
are affected. When settings are loaded, only those which are possible on the instrument
are implemented. Error messages indicate the settings which cannot be implemented.
The stored file is transferred from one instrument to another using the memory stick.
General file management functions such as copying and moving data are available in the
"File Manager" dialog.
5.2.8.1
File Menu
The settings available in the File menu "Save/Recall" depend on the operation selected
under "Select Operation".
For more information, see
chapter 4.7.1, "File Select Dialog"
Select Operation
Selects the file function.
General Instrument Settings