![Rohde & Schwarz CMW500 Series Скачать руководство пользователя страница 143](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718143.webp)
Remote Control
R&S
®
CMW500
143
User Manual 1173.9463.02 ─ 02
Manual and remote control
In contrast to mode other instrument settings, the repetition modes in manual and remote
control are independent and do not overwrite each other. The default repetition mode in
manual control is Continuous (observe results over an extended period of time), the
default mode in remote control is Single-Shot (perform one measurement and retrieve
results). See also
.
CONFigure:<FWA>:<Context>:REPetition <Count>
e.g.
CONFigure:GPRF:MEASurement<i>:POWer:REPetition CONT
Statistics Type
In general the following types of statistical results are available for scalar results (single
values) and traces:
Current
: Result of the current measurement interval
Minimum/Maximum
: Minimum or maximum of all evaluation periods since the mea-
surement was started
Average
: Average referenced to one single-shot measurement length
FETCh:<FWA>:<Context>:<View>:CURRent...?
READ:<FWA>:<Context>:<View>:CURRent...?
FETCh:<FWA>:<Context>:<View>:MINimum...?
READ:<FWA>:<Context>:<View>:MINimum...?
FETCh:<FWA>:<Context>:<View>:MAXimum...?
READ:<FWA>:<Context>:<View>:MAXimum...?
FETCh:<FWA>:<Context>:<View>:AVERage...?
READ:<FWA>:<Context>:<View>:AVERage...?
e.g.
FETCh:GPRF:MEAS<i>:POWer:CURRent:RMS?
Detector
Some measurements provide different detector settings to calculate the returned results
from the raw measurement data. An example is the GPRF Power measurement which
evaluates the maximum, minimum, and RMS (average) power within the current mea-
surement interval (see
). The detector is identified by means of an additional
mnemonic preceding the statistics type:
FETCh:<FWA>:<Context>:<View>:CURRent?
FETCh:<FWA>:<Context>:<View>:MINimum:CURRent?
FETCh:<FWA>:<Context>:<View>:MAXimum:CURRent?
READ:<FWA>:<Context>:<View>:CURRent?
READ:<FWA>:<Context>:<View>:MINimum:CURRent?
READ:<FWA>:<Context>:<View>:MAXimum:CURRent?
Control of the Instrument