Remote Control
R&S
®
CMW500
147
User Manual 1173.9463.02 ─ 02
FETCh:<FWA>:<Context>:TRACe:<View>:<Statistics>...?
READ:<FWA>:<Context>:TRACe:<View>:<Statistics>...?
e.g.
FETCh:GSM:MEASurement:MEValuation:EVMagnitude?
FETCh:GSM:MEASurement:MEValuation:TRACe:EVMagnitude:CURRent?
The
<View>
mnemonic is used in multi evaluation measurements where it distinguishes
different subsets of results. For detailed information about the returned values refer to
the command reference description. See also
.
6.4.2.5
Reliability Indicator
The reliability indicator is an information element that describes the validity of the results
and the possible source of inaccuracies or errors. The R&S CMW500 provides common
reliability indicators for an entire array of measurement results (retrieved with a single
command).
Common Reliability Indicator
The common reliability indicator is the first value in the output arrays of the
FETCh...?,
READ...?
and
CALCulate...?
queries.
Example:
FETCh:GPRF:MEAS:EPSensor:CURRent?
Response: 0, 10.22, 10.15, 10.01, 10.29, 100 (reliability = 0, followed by 5 numeric mea-
surement values).
The common reliability indicator is one of the following values:
●
0 (OK)
:
Measurement values available, no error detected.
●
1 (Measurement Timeout)
:
Measurement results may be available, however, at least a part of the measurement
provides only INValid results or has not completed the full statistic count. The mea-
surement has been stopped after the configurable measurement timeout.
●
3 (Overdriven) / 4 (Underdriven)
:
Measurement results available, however, the accuracy of some results may be
impaired because the input signal level was too high / too low.
●
6 (Trigger Timeout)
:
The measurement could not be started or continued because no trigger event was
detected.
●
7 (Acquisition Error)
:
The R&S CMW500 could not properly decode the RF input signal.
●
8 (Sync Error)
:
The R&S CMW500 could not synchronize to the RF input signal.
●
15 (Reference Frequency Error)
:
The instrument has been configured to use an external reference signal but the ref-
erence oscillator could not be phase locked to the external signal (e.g. signal level
too low, frequency out of range or reference signal not available at all).
●
16 (RF Not Available)
:
Control of the Instrument