![Rohde & Schwarz CMW500 Series Скачать руководство пользователя страница 205](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718205.webp)
Command Reference
R&S
®
CMW500
205
User Manual 1173.9463.02 ─ 02
Usage:
Query only
Firmware/Software:
V1.0.5.3
CONFigure:FDCorrection:ACTivate
<Connector>, '<TableName>'[, <Direction>]
Activates a correction table for the signal path using a specific RF connector.
For bidirectional connectors the table can be applied to both directions or to one direction.
It is possible to assign different tables to the directions of a bidirectional connector, see
example.
Depending on the installed hardware and the active sub-instrument only a subset of the
listed connector values is allowed. The mapping of virtual connector names to physical
connectors also depends on the active sub-instrument. For details see
Setting parameters:
<Connector>
RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAO | RFAC |
RFBC
RF1C, RF2C, RF3C, RF4C, RF1O, RF3O:
RF 1 COM to RF 4 COM and RF 1/3 OUT front panel connectors
RFAC, RFBC, RFAO:
Virtual names for the RF COM and RF OUT connectors
'<TableName>'
String parameter identifying the table. To display a list of existing
tables use the command
CONFigure:BASE:FDCorrection:CTABle:CATalog?
.
<Direction>
RXTX | RX | TX
Specifies the direction to which the correction table shall be
applied. RX means input and TX means output. For a pure output
connector RX is ignored.
RXTX:
both directions (for output connector only output)
RX:
input (no effect for output connector)
TX:
output
Default: RXTX
Example:
CONFigure:FDCorrection:ACTivate RF1C,
'mytable_in', RX
CONFigure:FDCorrection:ACTivate RF1C,
'mytable_out', TX
Different tables are activated for the input and output direction of
the connector RF1C.
Usage:
Setting only
Firmware/Software:
V1.0.5.3
CONFigure:FDCorrection:DEACtivate
<Connector>
Deactivates any correction tables for a specific RF connector.
Instrument-Control Commands