Remote Control
R&S
®
CMW500
173
User Manual 1173.9463.02 ─ 02
Each single status register in the figure above consists of five parts as described in section
Structure of an SCPI Status Register
. The following simplified presentation is used:
Fig. 6-2: Legend for figure above
The following table lists the assignment of network standards (and GPRF) to the bits of
the register STATus:OPERation:TASK:A. The mnemonics used to indicate the network
standards are identical to the mnemonics used in remote commands (see
Table 6-2: Bit definition for register STATus:OPERation:TASK:A
Bit No.
Network Standard
1
GPRF
2
GSM
3
WCDMa
4
WLAN
5
BLUetooth
6
WIMax
7
EVDO
8
CDMA
9
TDSCdma
10
LTE
The following table lists the assignment of the individual measurement and generator
applications to the bits of a function group register (STATus:OPERation:TASK:A:<net-
work standard>:MEASurement<i> / ...:GENerator<i>). The mnemonics for measure-
ments are identical to the mnemonics used in remote commands (see
). Generator function groups contain only one application which is not explicitly
Status Reporting System