RTD input (H-TIO-A/B/C/D/E/F/G/K/P/R, H-TI-A/B, H-CIO-A)
Input type
Range No.
0.0 to 400.0
°
C 59
0 to 400
°
C 38
−
200 to
+
200
°
C 39
−
200.0 to
+
200.0
°
C
58
JPt100
−
50.00 to
+
150.00
°
C
1
106
−
300 to
+
900
°
F 41
0 to 800
°
F 40
0.0 to 800.0
°
F 60
−
300.0 to
+
900.0
°
F
2
104
0.0 to 400.0
°
C 62
0 to 400
°
C 42
−
200 to
+
200
°
C 43
−
200.0 to
+
200.0
°
C
61
Pt100
−
50.00 to
+
150.00
°
C
1
107
−
300 to
+
1200
°
F 45
0 to 800
°
F 44
0.0 to 800.0
°
F 63
−
300.0 to
+
1200.0
°
F
2
105
Current input and Voltage input (H-TIO-H/J, H-CIO-A)
Input type
Range No.
Input group
0 to 10 mV DC
0.0 to 100.0 %
0
−
10 to
+
10 mV DC
0.0 to 100.0 %
1
0 to 100 mV DC
0.0 to 100.0 %
2
−
100 to
+
100 mV DC
0.0 to 100.0 %
3
Voltage (low) input group
Voltage
0 to 1 V DC
0.0 to 100.0 %
4
input
*
−
1 to
+
1 V DC
0.0 to 100.0 %
5
0 to 5 V DC
0.0 to 100.0 %
6
1 to 5 V DC
0.0 to 100.0 %
7
−
5 to
+
5 V DC
0.0 to 100.0 %
8
0 to 10 V DC
0.0 to 100.0 %
9
Voltage (high) input group
−
10 to
+
10 V DC
0.0 to 100.0 %
10
Current
0 to 20 mA DC
0.0 to 100.0 %
11
Current input group
input
*
4 to 20 mA DC
0.0 to 100.0 %
12
* Display scale of the current and voltage input can be changed.
An input type change may only be made within the input groups as shown above.
Pulse input (H-SIO-A)
Input
type
Range No.
Pulse
input
•
Dry contact input (Power supply for sensor,
12 V DC)
•
Voltage input (Power supply for sensor, 12 V DC)
Specify when ordering with model code.
0
Do not set any number other than 0, as this may cause malfunction.
1
The range with the resolution of 1/100 can be
specified only by H-TIO-E module.
2
The range can be specified only by H-TIO-F
module (high accuracy type).