5-1
IMSBR10-E1
Frequently Used Setup Operations (Setting Mode)
5
5.1 Setting the Input Range
Input range is set for each measurement channel. Set unused channels to Skip.
• The input range cannot be changed on models with the computation function (/M1 option) when computation
is in progress.
• If you change the input range, set the bias, alarm, and partial expanded recording again.
Procedure
TC, RTD, and DC Voltage
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key with
Range
shown on the screen.
3.
Set each item and press the
key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to then are
cancelled, and the display returns to a higher level menu.
4.
When the
Setting complete
screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the
ESC
key.
5.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Select TC, RTD, or Volt.
Set the channel range.
Select the range type.
Select the left span value.
Select the right span value.
The new setting takes effect.
ESC/?
Set=Range
First channel
Last channel
CH=01-01
Mode=TC
Range=R
Span left=
0.0
Span right= 1760.0
01-01 Channel
Setting complete
Chapter 5 Frequently Used Setup Operations (Setting Mode)