12-15
IMSBR10-E1
Specifications
12
Alarms
Number of alarms
Up to four alarms (level) for each measurement channels.
Alarm types
High limit (H), low limit (L), delay high limit (T), delay low limit (t). The alarm symbol is shown in parentheses.
Alarm delay time of delay high and low limit alarms (can be set for each channel, common within a channel) See
section 12.2
Hysteresis
Applied to high limit alarm and low limit alarm.
Hysteresis range: Approx. 0.0% to 1.0% of the recording span (0.1% steps)
Miscellaneous
See section 12.2.
Display
See section 12.4.
Cu10, Cu25 RTD Input (/N1)
Item
Specifications
Cu10, Cu25 RTD
This option allows Cu10 and Cu25 inputs to be added to the standard input types.
Input type and measurable range
Input Type
Measurable Range
RTD
(Measurement
current i: 2 mA)
Cu10 (GE)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10:
α
= 0.00392 at 20
°
C
Cu10:
α
= 0.00393 at 20
°
C
Cu25*:
α
= 0.00425 at 0
°
C
–200 to 300
°
C
–328.0 to 572.0
°
F
* Measurement current i = 1 mA
Measurement and recording accuracy
Cu10 (GE)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10:
α
= 0.00392 at 20
°
C
Cu10:
α
= 0.00393 at 20
°
C
Cu25:
α
= 0.00425 at 0
°
C
Measurement Accuracy
Recording Accuracy
Measurement accuracy
±
(0.3% of recording span)
±
(0.4% of rdg + 1.0
°
C)
±
(0.3% of rdg + 0.8
°
C)
Input Type
3 Legs Isolated RTD (/N2)
Item
Specifications
Input terminal isolation
Applies to the dot model.
The RTD input terminals are isolated between channels (on the standard dot model, input terminal b are common
to all channels)
Expansion Inputs (/N3)
Item
Specifications
Input type
Adds the input types shown in the table below.
Input
Type
Measurement
Range
PR40-20
0.0 to 1900.0
°
C
32 to 3452
°
F
PLATINEL
0.0 to 1400.0
°
C
32 to 2552
°
F
Thermocouple
NiNiMo
0.0 to 1310.0
°
C
32 to 2390
°
F
W/WRe26
0.0 to 2400.0
°
C
32 to 4352
°
F
Type N (AWG14)
0.0 to 1300.0
°
C
32 to 2372
°
F
Kp vs Au7Fe
0.0 to 300.0 K
Pt25
–200.0
to
550.0
°
C
–328.0 to 1022.0
°
F
Pt50
–200.0
to
600.0
°
C
–328.0 to 1112.0
°
F
Ni100 (SAMA)
–200.0 to 250.0
°
C
–328.0 to 482.0
°
F
RTD
Ni100 (DIN)
–60.0 to 180.0
°
C
–76.0 to 356.0
°
F
(Measuring current
Ni120
–70.0 to 200.0
°
C
–94.0 to 392.0
°
F
i = 1 mA)
J263*B
0.0 to 300.0 K
Cu53
–50.0 to 150.0
°
C
–58.0 to 302.0
°
F
Cu100:
α
= 0.00425 at 0
°
C
–50.0 to 150.0
°
C
–58.0 to 302.0
°
F
12.5 Specifications of Optional Functions