9-13
IMSBR10-E1
Operations Related to the Computation Function
(/M1 Option)
9
9.6 Specifying the Timer Used in Statistical
Calculations (TLOG)
On channels that perform statistical calculation (TLOG), the timer for determining the interval is specified. The sum
scale is also specified when using statistical calculation.
The timer is also used for the periodic printout of computed values.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key or
SHIFT
+
key to select
Math
and then press the
key.
3.
Press the
key or
SHIFT
+
key to select
TLOG
and then press the
key.
4.
Set each item and press the
key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to then are cancelled,
and the display returns to a higher level menu.
5.
When the
Setting complete
screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the
ESC
key.
6.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Set the channel range.
The new setting takes effect.
ESC/?
Select the timer number (Periodic, 1, or 2).
Select the sum scale.
First channel
Last channel
0A-0A Channel
Setting complete
Math=TLOG
Set=Math
CH=0A-0A
Timer number=Periodic
SUM scale=Off