2
Installation
2.1
Introduction
The C520 can be used as either a desk-top or panel-mount instrument. The C530 is intended for installation in
electronics cabinets. The C500 series contains precision electronics and must not be subject to shock, excessive
vibration, or extremes of temperature, either before or after installation. The operating environment must fall
within the allowed temperature range and humidity.
The inputs of the C500 series are protected against electrical interference, but excessive levels of electromagnetic
radiation and RFI may affect the accuracy and stability of the instrument. The C500 series should be installed
away from any sources of electrical noise. The loadcell cable is particularly sensitive to electrical noise, and
should be located well away from any power or switching circuits.
Termination of the loadcell shield at
the indicator end (with a sound connection to the indicator case via the DB9 plug or screw terminal shield
connection) is important for EMC immunity.
The C500 series must be installed in a manner and location that is protected from impacts.
Warning!
If the equipment is not installed and used as specified by the manufacturer, the
protection provided by the equipment may be impaired.
2.2
Environmental Conditions
The C500 series is designed for use in the following environmental conditions:
C520: Indoor use only. Although the front of the device can be washed down (as per IP65), the rear of
the device, along with cable attachments and optional accessories, must be protected from liquids and
small objects (as per IP30).
C530: Indoor use only. The device cannot be washed down. The device, along with cable attachments
and optional accessories, must be protected from liquids and small objects (as per IP30).
Altitude: up to 2000m
Operating temperature range: -10
°
C to 40
°
C
Storage temperature range: -20
°
C to 50
°
C
Humidity: Maximum relative humidity of 80% for temperatures up to 31
°
C decreasing linearly to 50%
at 40
°
C
Mains supply voltage fluctuations: up to
±
10%.
Pollution degree: 2
2.3
Mounting
2.3.1
C520 Panel Mounting
The C520 fits a DIN 43 700 standard 138(-0/+1)mm x 67(-0/+1)mm panel cut-out. Panel mounting is included
standard. To panel mount:
1. Insert the 4 supplied M4 screws in the mounting tabs in the corners.
2. Insert the indicator into the panel and rotate the 4 tabs out from the indicator.
3. Tighten the screws to clamp the indicator into the panel. Do not over tighten.
2
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...