21.3.39
POR: External IO
Force digital outputs on or off. This is only enabled for IO lines not configured as setpoints.
POR?
replies with
the status of the 8 digital IO lines.
General
No. of parameters
8
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1 - 8
Output 1 - 8
0 or 1
-
Example
Command
Answer
Description
S01;
Select unit 1
POR1,1,1,1,1,1,1,1;
0 CRLF
IO 1 - 8 set as outputs and set to on
TDD1;
0 CRLF
Save setting
LIV1,1;
0 CRLF
Setpoint 1 is active
POR1,1,1,1;
?
CRLF
Not possible as POR not enabled for setpoint 1
POR,1,1,1;
0 CRLF
Ok, only deactivated setpoint changed
POR?;
1,1,1,1,0,0,0,0 CRLF
IO 1 - 4 are on, IO 5 - 8 are off
21.3.40
PRS: Print and Automatic Output Settings
General
No. of parameters
10
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
148
C500-600-2.0.3
Содержание C500 series