Read Gross Weight
COMMAND A:
20050026:;
COMMAND A:
Read Gross Weight (Register 0026
H
):
ADDR
= 20
H
: Reply required from any instrument
CMD
= 05
H
: Read Literal
REG
= 0026
H
: Gross Weight
RESPONSE A:
81110026: 100 kg G;
RESPONSE A:
Same response from instrument #1 but in literal format.
Zero calibrate scale
COMMAND A:
21100102:;
COMMAND A:
Attempt to zero calibrate scale (Register 0102
H
):
ADDR
= 21
H
: Reply required from instrument #1
CMD
= 10
H
: Execute
REG
= 0102
H
: Zero calibrate scale
RESPONSE A:
C1100102:0601;
RESPONSE A:
Instrument #1 reports “ERROR: Exec Error, Permission error”.
(Writing to this register requires a passcode).
COMMAND B:
21170019:1234;
COMMAND B:
Enter full passcode (Register 0019
H
):
ADDR
= 21
H
: Reply required from instrument #1
CMD
= 17
H
: Write Literal Decimal
REG
= 0019
H
: Gross Weight
DATA
= 1234: Example passcode
RESPONSE B:
81170019:0000;
RESPONSE B:
Instrument #1 reports “Passcode Accepted”.
COMMAND C:
21100102:;
COMMAND C:
Zero calibrate scale (Register 0102
H
):
ADDR
= 21
H
: Reply required from instrument #1
CMD
= 10
H
: Execute
REG
= 0102
H
: Zero calibrate scale
RESPONSE C:
81100102:0000;
RESPONSE C:
Instrument #1 reports successful calibration.
COMMAND D:
21100010:;
COMMAND D:
Save the settings (Register 0010
H
):
ADDR
= 21
H
: Reply required from instrument #1
CMD
= 10
H
: Execute
REG
= 0010
H
: Save settings
RESPONSE D:
81100010:0000;
RESPONSE D:
The settings have been successfully saved.
76
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...