Note
The hexadecimal codes are combined in the fields described above when multiple
options are active at the same time. For example an error response message from
instrument address 5 would have an
ADDR
code of C5
H
(80
H
+ 40
H
+ 05
H
).
18.2.2
Termination
Message termination is possible in two ways. For normal communications that do not involve checksums use
either a
CR LF
(ASCII 13, ASCII 10) as a terminator or a semicolon (‘;’ ASCII). There is no start-of-message
delimiter.
To use a checksum the message is framed as:
SOH <Message> CRC EOT
where:
SOH (1 byte): 01
H
CRC is a 4 character hexadecimal field comprising the 16 bit CRC checksum.
The CRC uses the
16 bit CCITT polynomial calculation and includes only the contents of the “Message” section of the
transmission.
EOT (1 byte): 04
H
18.2.3
Error Handling
If a command cannot be processed and a reply was requested, the ERROR bit in the ADDR field is set and
the DATA field is set to an error message. This 2-byte error message is formed by adding a major and minor
error code. For example, the error 0204
H
corresponds to a major error of 0200
H
, and a minor error of 04
H
).
Error Code
Error
Description
0100
H
Parsing Error
The error occurred when interpretting the incoming message.
0101
H
Address error
An error occurred when parsing
ADDR
into a number.
0102
H
Command error
An error occurred when parsing
CMD
into a number.
0103
H
Reg error
An error occurred when parsing
REG
into a number.
0104
H
Chan error
The internal channel string is longer than the maximum (9).
0105
H
Delimiter error
The colon delimator is not present.
0106
H
Data error
The data string is longer than the maximum (200)
0108
H
Framing error
The message is too long for the buffer.
0200
H
CRC Error
There was a problem with the checksum (CRC).
0201
H
CRC Parse Error
The CRC code could not be parsed from the message.
0202
H
CRC Error
The CRC code does not match that of the message. The message
was corrupted.
0300
H
Register Error
REG
is not valid, or does not exist on this device.
0400
H
Read Error
There was an error reading from the register.
0401
H
Permission error
You do not have permission to read this register.
0402
H
Unknown type error
Internal error: Register type is not known to RinCmd.
0403
H
No type data error
Internal error: Unable to read the registers data structure.
0404
H
Command error
The
CMD
is not valid for this register type.
0405
H
Bad data error
The stream data is not valid.
0406
H
Null register error
The register cannot be found.
0407
H
Buffer size error
The string reply will not fit in the transmission buffer.
0408
H
Snprintf error
A function within the library has failed to run correctly.
74
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...