21.3.9
BUZ: Buzzer
Enable/disable the buzzer.
General
No. of parameters
1
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1
Enable
0 = Off
1 = On
1
Example
Command
Answer
Description
S01;
Select unit 1
BUZ?;
1 CRLF
Query setting, buzzer currently enabled
BUZ0;
0 CRLF
Buzzer disabled
TDD1;
0 CRLF
Save new settings
21.3.10
CDL: Zero
Set the zero dead load cancellation. This is analogous with pressing the Zero key on the front of the instrument.
General
No. of parameters
0
Save changes
At input
Increment Trade Counter
No
Example: Set zero dead load of unit 1
Command
Answer
Description
S01;
Select unit 1
CDL;
0 CRLF
Zero dead load set successfully
<
load disturbed
>
CDL;
1 CRLF
Scale in motion
Table of possible replies:
?
CRLF
Command not understood
0 CRLF
Command correctly operated
1 CRLF
Scale in motion
2 CRLF
Range out. For example, zero setting out of range
3 CRLF
System error
4 CRLF
Device not ready. For example, printer not ready
21.3.11
CFM: Clock formats
Clock time and date formats.
General
No. of parameters
2
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
127
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...