The indicator contains both current and latched error status flags. The latched errors are only cleared by
resetting the unit (
RES
command or power off). The response string is 5 hexadecimal characters representing
the 20 error bits. See Section 24.4.
Example
Command
Answer
Description
S01;
Select unit 1
ESR?;
00000 CRLF
No current errors
ESR?1;
000C0 CRLF
Positive and negative sense lines were not connected at some time in the past.
21.3.19
ETH: Ethernet
Set the Ethernet port parameters - Internet Protocol (IP) address, network mask, etc.
General
No. of parameters
10
Save changes
ETH; command will save changes
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1
DHCP Enable
0 .. 1
1
2
IP address
0.0.0.0 - 255.255.255.255
192.168.1.254
3
Subnet mask
0.0.0.0 - 255.255.255.255
255.255.255.0
4
Gateway address
0.0.0.0 - 255.255.255.255
192.168.1.1
5
DNS1 address
0.0.0.0 - 255.255.255.255
192.168.1.1
6
DNS2 address
0.0.0.0 - 255.255.255.255
0.0.0.0
7
MAC address
(read only)
8
Hostname
1 .. 64 characters from a-z, 0-9, -
c520-
<
serial number
>
or
c530-
<
serial number
>
9
Search domain
0 .. 255 characters from a-z, 0-9, -, .
“”
10
DNS3 address
0.0.0.0 - 255.255.255.255
0.0.0.0
Note
Setting the ETH command over the network may result in a lost connection if the
IP address changes as a result. If this occurs, use the device finder to re-locate
the indicator and re-connect. Alternatively, use the rinlink port to change these
parameters.
Example
Command
Answer
Description
S01;
Select unit 1
ETH?;
0,"192.168.1.100","255.255.255.0","192.168.1.1",
"192.168.1.1","0.0.0.0","70:4A:E4:00:00:01",
"c520-3382121","search.domain","0.0.0.0"CRLF
Read Ethernet settings
(static)
ETH1;
0 CRLF
Enable DHCP
ETH?;
1,"172.17.1.110","255.255.255.0","172.17.1.1",
"172.17.1.2","172.17.1.3","70:4A:E4:00:00:01",
"c520-3382121","new.search.domain","0.0.0.0"CRLF
Read Ethernet settings
(set by DHCP)
21.3.20
FBT: Front Function Keys
Setup the functions of each of the 3 front function keys or artificially force the execution of the function.
133
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...