Parameter
Description
Range
Default
1
Mode of operation
0 = Normal print
1 = Perform displayed weight printout, and reply with print
ID, date, time and displayed weight
2 = Same as 1, but using gross weight
3 = Same as 1, but net weight
4 = Same as 1, but tare weight
0
2
String to be printed
Up to 250 characters
No escaping required for printable ASCII characters
(see Section 9.5)
Escape non-printable ASCII characters using a back-
slash followed by a three digit number (see Section 9.5)
Escape printing tokens using a backslash followed by
a letter (see Section 9.4)
-
Table of possible replies:
?
CRLF
Command not understood
0 CRLF
Command correctly operated
1 CRLF
Scale in motion
2 CRLF
Range out. For example, zero setting out of range
3 CRLF
System error
4 CRLF
Device not ready. For example, printer not ready
Query Parameter Details
Parameter
Description
Range
1
Data selection
0 = ID of last printout
1 = Next line of printout as a string in quotation marks.
For example, “ WEIGHT ”. A 2048 byte buffer is used to
store printouts. This command returns a single line of a
printout. Where lines longer than 100 characters, they are
cut into pieces. Each line is discarded after it is sent. An
empty string is sent when the buffer is empty.
Note
Poll PRT?0 to determine when new data is available, then use PRT?1 for each line
of the printout until an empty string is returned. Non-printable characters are sent
using
\
xxx format. For example,
CRLF
is
"
\
013
\
010"
.
Example
150
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...