No. of parameters
6
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1
Jog on time (milliseconds)
100 .. 60000
100
2
Jog off time (milliseconds)
100 .. 60000
100
3
Jogs in set
1 .. 20
1
4
Max jog sets
0 .. 20
0
5
Feeder type
0 = MULT.
1 = SINGLE
0
6
Delay check (milliseconds)
0 .. 60000
0
Example
Command
Answer
Description
S01;
Select unit 1
LIW?;
100,100,1,0,0,0 CRLF
Query general setpoint parameters
LIW200,100,5,0,0,0;
0 CRLF
Change to: 200ms jog on, 100ms jog off, 5 jogs in a set,
unlimited sets, multiple feeders, no check delay.
TDD1;
0 CRLF
Save setting
21.3.31
LWN: Span Calibration
Calibrate the span of the scale. This is the new form of the
LWT;
command (see Section 21.3.33). Because
the C500 series has no concept of normal and direct calibration mode, it was necessary to introduce a new
command.
Note
There are two different methods of calibration. Either calibrate with standard
weight or direct input of mV/V.
General
No. of parameters
2
Save changes
With
TDD1
Increment Trade Counter
Yes
Parameter Details
Parameter
Description
Range
Default
1
Span signal in deci-micro volts
per volt.
10000duV/V = 1000uV/V =
1 mV/V
-50000 .. 50000
-
2
Status (see table below)
0 .. 106
-
Calibration Status
Status value
Description
0
Calibration finished successfully
1
Calibration in process (busy)
104
Error span too high (
>
5.0 mV/V) or too low (
<
-5.0 mV/V). Calibration aborted.
106
Other calibration error. Calibration aborted.
Example: Normal Calibration with Weight
142
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...