90BRotary pump (2080A211/A411)
SmartLab: Automated Multipurpose X-ray Diffractometer
105
Table Patterns of the OPERATE lamp flashing and alarm sounds when an alarm generated
Alarm point
Color
Continuously
On / flashing
Pattern
XG alarm
Green
Flashing
One long beep and two short beeps (repeats)
Attachment
connecting
cover is open
Green
Flashing
Two short beeps (repeats)
Communication
error
Red Flashing
One beeps (repeats)
Encoder error
Red
Flashing
Two beeps (repeats)
Servo alarm
Red
Flashing
Three beeps (repeats)
Table Causes of alarm
Cause
Description
XG alarm
An error has occurred in the x-ray generator (XG). See # 4 in
Attachment
connecting cover is
open
The cover of the attachment connection cable is open in the radiation
enclosure. Close the cover.
Communication
error
A communication error has occurred in the control board for the optical
devices in the incident and receiving optics. The cables for the incident and
receiving optics may be disconnected or broken.
Encoder error
An error has occurred in the encoder used to control the goniometer. The
encoder may not be adjusted correctly, or the detecting part may be stained.
Servo alarm
The protection circuit is working as a result of motor electrical overload. See
400 ms
1000 ms
50 ms
200 ms
1000 ms
1000 ms
1000 ms
50 ms
50 ms
50 ms
200 ms
200 ms
200 ms
200 ms
200 ms
200 ms
200 ms
2500 ms