Shutdown
SmartLab: Automated Multipurpose X-ray Diffractometer
21
Shutdown
Click
Package Measurement
under the
Task
menu and click
Shutdown
in the
displayed
Package Measurement
flow bar.
Shutdown
CAUTION
When not using the water circulation pump, wait three minutes following
shutdown (i.e., three minutes after the yellow lamp located at the upper
section of SmartLab goes out), then halt the supply of cooling water. (Halting
the cooling water supply in less than three minutes may result in damage to
SmartLab.)
For more information on the shutdown procedure, refer to Section 2.6 of the “CBO
Maintenance Package Measurement” Help Topic of the online help section of the
SmartLab
Guidance software
.
While SmartLab is not used, it is recommended to turn off only the x-ray, and do not turn
off the vacuum unit. In this case, the controller remains powered; therefore SmartLab can be
started from SmartLab Guidance. If SmartLab is unused for a while (e.g. during the New
Year Holidays), press the “O” OFF button to turn the power of SmartLab off.