Receiving optics unit # 1
SmartLab: Automated Multipurpose X-ray Diffractometer
51
Table Parallel slit analyzers
Aperture angle
Length
Abbreviation
Illustration
Open 45
mm
PSA_open
1.0º
(option: 2680F113)
45 mm
PSA_1.0 deg
0.5º
45 mm
PSA_0.5 deg
0.114º
(option: 2680F111)
90 mm
PSA_0.114 deg
Open 90
mm
PSA_open
The standard configuration of SmartLab cannot incorporate either the PSA 0.114
deg. or PSA open (90 mm) in conjunction with a receiving parallel slit. In order to
do this, the configuration must be modified by, moving the receiving parallel slit
adaptor, standard slit box # 2, attenuator, and detector (counter) to the right.
The type of parallel slit analyzer installed on the adaptor can be identified from SmartLab
Guidance.