Detector
72
SmartLab: Automated Multipurpose X-ray Diffractometer
2
Install the unit on the detector adaptor. Secure it in place with the
Allen wrench provided.
3
The unit installation position can be read from the scale on the theta_d
arm.
When using the monochromator in Bent mode with para-focusing optics, the unit
must be installed at a specific position for a specific target given in the table below.
When using the monochromator in Flat mode, the exact position is not important.
Type of target
Installation position
Cu
351.5 mm
Co
359.8 mm
Fe
364.7 mm
Cr
376.1 mm