Receiving slit box # 1
SmartLab: Automated Multipurpose X-ray Diffractometer
47
2
Connect the cable to the SLIT1 connector on the receiving connector
box.
3
A
K
β
filter may be inserted.
Check whether a K
β
filter is inserted from SmartLab Guidance.
K
β
filter (Ni foil)
The K
β
filter reduces the K
β
characteristic x-rays by approximately 90% while only
reducing the K
a
characteristic x-rays by approximately 50%. Therefore almost all K
β
x-rays
can be removed. (Used for measurements based on the para-focusing method.)