SmartLab: Automated Multipurpose X-ray Diffractometer
79
Sample Stage
To move the sample up and down,
Z
axis is installed on the SmartLab goniometer.
This mechanism enables the sample surface to be adjusted when using a sample
stage or an attachment.
Sample height alignment unit
Sample stages can be installed to the sample height alignment unit. To install a
sample stage, join the stage with the rail of the sample height alignment unit as
shown in the following figure, and slide it from top down. Make sure that the stage
is placed horizontally and secure it in place with the Allen wrench provided.
The direct beam stopper (option: provided with 5741A300) and knife edge (option:
2431G101) can also be installed to the sample height alignment unit. They will
reduce the background intensity in high-speed measurements using D/teX Ultra.
Direct beam stopper