SmartLab: Automated Multipurpose X-ray Diffractometer
85
Attachments
You can switch between attachments with various adjust axes for different
measurement purposes.
Described below is the method for changing attachments.
1
Click Manual Control under the Control menu. Move the chi axis to 0
°
and the phi axis to 0
°
.
2
Release the hook.
3
Open the clasps for the right and left sides. Remove the attachment.
4
Make sure the right and left clasps are open.
5
Install the attachment by aligning the groove of the attachment with
the pin on the front.
6
Close the clasps and engage the front hook.
Pin
Hook
Clasps