SmartLab: Automated Multipurpose X-ray Diffractometer
23
Opening and Closing the Door
To prevent human exposure to x-ray radiation, the shutter and door are equipped
with an interlocking function. To ensure safety, when changing samples or optical
devices during x-ray generation, open and close the door by the following
procedure:
Opening the door
Under normal conditions, the door is locked and cannot be opened.
1
Press the DOOR LOCK button on the door.
2
The shutter will close and the door will be unlocked.
3
The DOOR LOCK button will flash, and you will hear beeping.
4
Open the door.
The shutter cannot be opened when the door is unlocked. The following message
also appears when the door is unlocked. SmartLab Guidance cannot be operated
until the door is locked again.