- 50 -
optional moment and saving records from the carried out tests in the form of printed reports of the
tests that are automatically generated at the end of examination. Up to 50 reports can be recorded.
The test controls repeatability of placing the interval weight and determining error of indication with
reference to balance’s maximum capacity.
Testing procedure:
•
deposit internal weight twice,
•
deposit internal weight ten times, calculate the value of standard deviation,
•
perform balance adjustment,
•
print a report.
Test results provide balance data, calculated error for Max capacity and value of repeatability of
indication expressed as standard deviation.
An exemplary report:
.....................................................
--------- Autotest GLP: Report ---------
Balance type PS 3000.R2
Balance ID 400010
User Admin
Software rev. v.0.4.9
Date 2013.07.16
Time 09:17:16
----------------------------------------
Number of measurements 10
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
Filter Average
Value release Fast &Reliable
----------------------------------------
Deviation for Max. -0.118 g
Repeatability 0.0088 g
Signature
........................................
Procedure:
Enter <OTHER> parameters’ group, start <AUTOTEST GLP> parameter.
Press <Start> field. Autotest GLP procedure starts, it is carried out automatically. Autotest procedure
progress bar is displayed.
Press
or
button to abort the procedure, you can do it at any time.