- 109 -
3. The system detects the new USB device and automatically starts searching a respective driver
4.
Go to Device Manager and check number assigned to
virtual COM port.
For this very case it is STMicroelectronics Virtual
COM Port (COM8).
5. Set balance parameters: select USB value for parameter COMPUTER/PORT.
6. Run software for measurement readout.
7. Use the program to set parameters for communication – select respective COM port (for the
very case it is COM8), i.e. the one which has been formed while installing drivers.
8. Start cooperation.
25. PERIPHERAL DEVICES
PERIPHERAL DEVICES menu is comprised within Parameters menu. It is accessed by pressing
key. The menu features list of devices that can cooperate with the balance.
25.1. COMPUTER
Carry out configuration using <Computer> submenu.
Procedure:
•
press
key, next press: <Peripherals / Computer>
•
set parameters related to cooperation with a computer
•
computer port
available options: none, COM 1, COM 2, USB, Ethernet, Wi-Fi
CAUTION: Communication via Ethernet does not refer to WLC X2 series.
•
port settings
settings related to a port selected for computer connection
CAUTION: for Ethernet and Wi-Fi port Timeout parameter is on. The parameter specifies
time delay – time interval for disconnection, counted from the moment of the most recent
command sent from the device to which the balance is connected. Timeout parameter
value has to be entered in milliseconds [ms].
•
E2R
options: Yes/No
E2R is a control system designed to monitor all weighing processes carried out on a
balance, with E2R system activated some databases can be operated by computer
exclusively (access to specified functions and options using balance is not possible).
To start cooperation with E2R system, set <E2R> parameter to <YES> value, as a
result a respective pictogram is displayed in the top bar: