- 74 -
•
Remove the sample form the weighing pan.
•
All samples weighed afterwards are compared with the reference mass, and the display
indicates the difference, expressed in [%], for each weighed sample in relation to the reference
mass.
PROCEDURE 3
– selecting product from Products Database
•
Press <
Product> button, using displayed product list select the product that is to be
weighed.
•
,,Info” field automatically displays new information on reference mass and product.
•
Reference mass relating to the product is recorded automatically and displayed in reference
mass field.
•
0.000 % indication is displayed (for unloaded pan).
•
All samples weighed afterwards are compared with the reference mass, and the display
indicates the difference, expressed in [%], for each weighed sample in relation to the reference
mass
17. DENSITY OF SOLIDS
<
Solids density>
is a working mode enabling determination of density of a representative
sample material.
In order to operate this mode, an optional density kit (supplementary equipment), appropriate to
a model of used balance, is required. Prior to installing the kit it is necessary to remove the weighing
pan and a draft shield.