User Manual 3153
Specifications
A-8
Channels 2 and 3 High Level Output
Harmonics and non-related spurious
< -55dBc for carrier frequencies < 1MHz
< -40dBc for carrier frequencies < 5 MHz
In-band spurious and non-harmonic at 20MHz. Measured with sample clock = 50 MHz, 400-cycle sine
waveforms in a 1000 point segment
< -25dBc for amplitude
≤
5Vp-p
< -15dBc for amplitude
≤
10Vp-p
Two Tone inter-modulation
< -35dBc for 12.1MHz & 13.1MHz,
≤
5Vp-p
< -30dBc for 23.1MHz & 24.1MHz,
≤
10Vp-p
Channels 2 and 3 Low Level Output
Harmonics and non-related spurious
< -60dBc for carrier frequencies < 1MHz
< -45dBc for carrier frequencies < 5 MHz
In-band spurious and non-harmonic at 20MHz. Measured with sample clock = 50 MHz, 400-cycle sine
waveforms in a 1000 point segment
< -35dBc
Two Tone inter-modulation
< -55dBc for 12.1MHz & 13.1MHz