User Manual 3153
Maintenance and Performance Checks 5-16
Required
Equipment
Recommended equipment for calibration is listed below. Instruments
other than those listed may be used only if their specifications equal or
exceed the required minimal characteristics. Also listed below are
accessories required for calibration.
SMB to BNC cable
Dual banana to BNC adapter
50
Ω
Feedthrough termination
Equipment
Model No.
Manufacturer
Digital Multimeter
2000
Keithley
Adjustment
Procedures
Use the following procedures to calibrate the Model 3153. The following
paragraphs show how to set up the instrument for the calibration and
what are the acceptable limitations for the calibrated function.
Figure 5-6, Left Side 3153
RV6
RV1
RV2
RV4
RV3
RV5
RV7