User Manual 3153
Maintenance and Performance Checks 5-12
4. Set sequence advance mode to single or all three (3) channels.
(INST:SEL 1|2|3;:SEQ:ADV SING)
5. Perform sequence test using Table 5-7c. Each trigger advances the
sequence to the next step.
Table 5-7c, Sequence Operation – Single Advance.
Step
External Trigger Signal
Oscilloscope
Reading
Ch 1
Pass Fail
Ch 2
Pass Fail
Ch 3
Pass Fail
1
No trigger
Segment 1
2
20ns width, 2 shot
Segment 2
3
20ns width, 2 shot
Segment 3
4
20ns width, 2 shot
Segment 4
5
20ns width, 2 shot
Segment 5
6
20ns width, 2 shot
Segment 1
Mixed Advance
1. Configure model 3153 to output as was done in the Step Sequence
Advance test. Set sequence advance mode to mixed.
2. Program steps 2 and 3 with the mixed mode bit = 1. Use WaveCAD to
modify the sequence table and download on all three (3) channels.
3. Perform sequence test using Table 5-7d.
Table 5-7d, Sequence Operation – Single Advance.
Step
External Trigger Signal
Oscilloscope
Reading
Ch 1
Pass Fail
Ch 2
Pass Fail
Ch 3
Pass Fail
1
No trigger
Segment 1
2
20ns width, 2 shot
Segment 2
3
20ns width, 2 shot
Segment 3
4
20ns width, 1 shot
Segment 4
5
20ns width, 1 shot
Segment 5
6
20ns width, 1 shot
Segment 1