User Manual 3153
SCPI Command Reference 4-5
Boolean Parameters
Boolean parameters represent a single binary condition that is either
true or false. The generator accepts "OFF" or "0" for a false condition.
The generator accepts "ON" or "1" for a true condition. The instrument
always returns "0" or "1" when a boolean setting is queried. The
following command uses a boolean parameter:
OUTP:FILT { OFF | ON }
The same command can also be written as follows:
OUTP:FILT {0 | 1 }
Arbitrary Block
Parameters
Arbitrary block parameters are used for loading waveforms into the
generator's memory. Depending on which option is installed, the Model
3153 can accept binary blocks up to 1M bytes. The following command
uses an arbitrary block parameter that is loaded as binary data:
TRAC:DATA#564000<binary_block>
SCPI Command
Summary
Table 4-1 summarizes the complete SCPI command tree available to
program the generator over the GPIB. Refer to earlier sections in this
manual for more complete details on each command.
Table 4-1. Model 3153 SCPI Command Summary
Keyword
Parameter Form (Default in Bold)
SCPI 1993.0
Notes
:FORMat
:BORDer
NORMal | SWAPped
Confirmed
:WAVE
NORMal | USER
Not confirmed
:INSTRument
Confirmed
:SELect
1 | 2 | 3
Confirmed
:COUPLe
NONE | 2 | 3 | ALL
Confirmed
:PHASe
Confirmed
[:OFFSet]
(0;0;n)
Confirmed
n = waveform length
:OUTPut
Confirmed
[:STATe]
OFF | ON
Confirmed
:TYPE
HIGH | LOW
Confirmed
:FILTer
Confirmed
[:LPASs]
NONE | 1 | 2 | ALL
Confirmed
:TTLTrg<n>
(0;0;3)
Confirmed
(default;min;max)
[:STATe]
OFF | ON
Confirmed
:SYNC
Not confirmed