Specifications A-9
Outputs (Each Channel)
Main Output
Connector
Front panel SMB
Standby
Output Off or Normal
Impedance
50
Ω
, ±1%
Protection
Protected against short to case ground
Amplitude
Range
10mV to 10V
pk-pk
into 50
Ω
in three internal ranges
Resolution
3.5 digits, each range
Accuracy (1kHz)
±1%
Offset
Range
0 to ±4.5 V in three internal ranges, amplitude
dependent
Resolution
3.5 digits, each range
Accuracy
±1%
Amplitude-Offset Interaction
Description
Offset voltage is injected at the input of the output amplifier.
Therefore, reducing the output amplitude has an automatic
affect on the offset voltage. The 3153 has three internal
amplitude ranges that are transparent to the user.
However, it is necessary to understand these ranges in
case a specific amplitude-offset combination is needed.
The relationship between the internal amplitude ranges,
their respective offset ranges and the resultant peak range
limits are given in the equation and the table below.
imit
PeakRangeL
Offset
Amplitude
≤
+
2
Internal Amplitude Range
Offset Range
Peak Range Limit
10.00mV to 99.99mV
0 to 45.00mV
50mV
100.0mV to 999.9mV
0 to 450.0mV
500mV
1.000V to 10.00V
0 to 4.500mV
5V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com